Product Details:
|
Wavelength Resolution: | 1.0 Pm | Absolute Wavelength Accuracy: | ±10 Pm, Typ. ±5 Pm |
---|---|---|---|
Power Stability: | ±0.005dB (15Min) ±0.05 DB ( 24 Hours) | Power Linearity: | ± 0.3 DB |
Power Measuring Range: | +5 DBm To -85 DBm | Operating Temperature: | 0°C ~ +45°C |
Storage Temperature: | −40°C ~ +80°C | ||
Highlight: | Two Channels Laser Scanning Spectrometer,Constant Temperature Laser Scanning Spectrometer |
Two Channels Laser Scanning Spectrometer ±5 Pm Typ., 24 Hrs At Constant Temperature
The U8512 laser scanning spectrum analyzer is a passive device testing and spectrum analysis system controlled by PC. It is mainly used for coarse WDM components, high-density WDM components, Interleaver, AWG, WSS, OPM, DPSK components and other general optical fiber optical measurement and application. The tunable laser light source is integrated in the system, and with the optical power meter, it can be a tunable laser light source test system platform with high wavelength accuracy and fast scanning function.
Specification
Model # |
UC8512C+UC88201 |
Adjustable laser light source | |
Wavelength range |
1525.00 ~ 1568.00 nm |
Output Power |
13 dBm |
Wavelength resolution |
1.0 pm |
Absolute wavelength accuracy |
±10 pm, Typ. ±5 pm |
Relative wavelength accuracy |
± 5 pm, Typ. ±2 pm |
Wavelength repeatability |
± 3 pm, typ. ±1 pm |
Wavelength stability |
±5 pm (typ., 24 hrs at Constant Temperature) |
Tuning speed | ≤ 2 ms per step |
Power stability |
±0.005dB (15Min) ±0.05 dB ( 24 hours) |
Power repeatability |
± 0.05 dB |
Power linearity |
± 0.3 dB |
Power flatness | 0.3 dB; Typical values 0.05dB. |
Side mode rejection rate | ≥ 40 dBm |
Relative noise intensity | Typical values < -135 dB |
Optical power meter | |
Wavelength range | 850 nm to 1625 nm |
Power measuring range | +5 dBm to -85 dBm |
Precision of power | < ± 4% |
Power repeatability | < 0.1 dB |
Scanning function: equipped with DWDM/Isolator/GFF device scanning software | |
Scanning mode |
Fast adjustable laser spectral scanning |
resolution ratio | 1 pm |
precision | 5 pm |
Power display | Peak power display |
Central wavelength function | Central wavelength and 3dB bandwidth display |
Reference function | Reference test |
Power adapter power | AC 100 ~ 240 V , 48 - 66 Hz,100 VA max. |
display | PC control interface |
Optical output interface | FC/PC connector |
Communication interface | USB / RS232 interface |
Storage temperature | −40°C ~ +80°C |
Operating temperature | 0°C ~ +45°C |
Dimensions | 245 mm W, 105 mm H, 320 mm D |
Weight | 3.0kg |
The power supply requirements
The UC8512C+UC88201 complies with overvoltage Level II regulations. The operating voltage is 100 ~ 240V AC voltage, the frequency range is 48 ~ 66Hz, the operating voltage is 115V, the maximum consumption current is 230mA; When the operating voltage is 230V, the maximum consumption current is 120mA.
USB port
The USB interface is a standard 4-core B-type interface, which can be connected to UC8512C+UC88201 and PC, and the instrument can be controlled by monitoring software.
USB connection cable
The length of the USB cable cannot exceed 5 m. Otherwise, only a third-party USB peripheral extender can be used. Generally, the maximum extension length can reach 50 m.
RS232 serial port
The serial port parameters of the instrument are set to a fixed value. The serial port parameter Settings of the computer must match the instrument parameter Settings.
RS232 interface pin definition
Contact Person: Jack Zhou
Tel: 13602867834
Fax: 86-020-82575318